arXiv:2603.08983cs.ROcs.CV2026-03被引 1

无需标记点,用高斯点云实现达芬奇手术机器人手眼标定

SurgCalib: Gaussian Splatting-Based Hand-Eye Calibration for Robot-Assisted Minimally Invasive Surgery

  • 基于高斯溅射的可微渲染,结合运动约束优化器械位姿
  • 左右机械臂工具尖端定位误差分别低至4.75毫米和5.98毫米
  • 适合无菌环境下自动标定,避免增加手术室额外标记物

本文提出SurgCalib,一种基于高斯溅射的达芬奇手术机器人手眼标定框架。在视觉引导的机器人系统中,准确估计机器人基座与相机坐标系间的刚性变换对闭环控制至关重要。对于电缆驱动的手术机器人,编码器因缆绳拉伸和回程误差导致本体感觉测量不准,带来独特挑战。传统手眼标定依赖已知的标定图案并求解AX = XB方程,但引入额外标记物可能破坏无菌环境且干扰手术流程。为此,我们提出SurgCalib——一种全自动、无标记的标定方法。该方法先用原始运动学数据初始化器械姿态,再在高斯溅射的可微渲染框架下,结合遥控中心(RCM)约束进行两阶段优化。我们在公开的dVRK基准数据集SurgPose上进行了评估,结果显示左右机械臂工具尖端的平均2D重投影误差分别为12.24像素(2.06毫米)和11.33像素(1.9毫米),3D欧氏距离误差分别为5.98毫米和4.75毫米。

原文摘要 · Abstract (English)

We present a Gaussian Splatting-based framework for hand-eye calibration of the da Vinci surgical robot. In a vision-guided robotic system, accurate estimation of the rigid transformation between the robot base and the camera frame is essential for reliable closed-loop control. For cable-driven surgical robots, this task faces unique challenges. The encoders of surgical instruments often produce inaccurate proprioceptive measurements due to cable stretch and backlash. Conventional hand-eye calibration approaches typically rely on known fiducial patterns and solve the AX = XB formulation. While effective, introducing additional markers into the operating room (OR) environment can violate sterility protocols and disrupt surgical workflows. In this study, we propose SurgCalib, an automatic, markerless framework that has the potential to be used in the OR. SurgCalib first initializes the pose of the surgical instrument using raw kinematic measurements and subsequently refines this pose through a two-phase optimization procedure under the RCM constraint within a Gaussian Splatting-based differentiable rendering pipeline. We evaluate the proposed method on the public dVRK benchmark, SurgPose. The results demonstrate average 2D tool-tip reprojection errors of 12.24 px (2.06 mm) and 11.33 px (1.9 mm), and 3D tool-tip Euclidean distance errors of 5.98 mm and 4.75 mm, for the left and right instruments, respectively.

手术机器人手眼标定高斯溅射无标记

Thank you to arXiv for use of its open access interoperability. PaperDance 不是 arXiv 官方产品;中文卡片由大模型生成,请以原文为准。